Measure the conductivity after the reverse osmosis membrane
Measure the conductivity of the tap water
Do the following calculation : 100 * (conductivity reverse osmosis) / (conductivity raw water)
If the value is less than 10%, the membrane is still usable
In the other case, the reverse osmosis membrane begins to saturate.