- verify the main power is active - RED Light on the power switch button must be ON.

- verify the 24V output of the adaptor. Replace the adaptor if necessary.

- if 24VDC is OK; verify the fuse inside the controller box. Before opening, switch off first the system.

allindicatoroff.1552903293.txt.gz · Last modified: 2019/03/18 10:01 by thierry