**This manual process is similar to the flush function available on the system equipped with an electronic controller, which allows to rince the membrane easily just by pressing a button.** | **This manual process is similar to the flush function available on the system equipped with an electronic controller, which allows to rince the membrane easily just by pressing a button.** | ||
</WRAP> | </WRAP> | ||
+ | |||
+ | ===== Troubleshooting ===== | ||
+ | |**PROBLEM**|**CONDITION**|**POTENTIAL CAUSES **|**SUGGESTED ACTION **| | ||
+ | | Water purifier OFF| Power light OFF|No power supply|- Verify the power supply wires \\ - Switch power ON | | ||
+ | | ::: | ::: | Power supply fuse is broken| Check and replace the fuse| | ||
+ | |No production of purified water|- Power light ON \\ - indicator PUMP is off \\ - indicator SOURCE is ON| The inlet tap water supply is closed or has insufficient pressure|Improve water supply| | ||
+ | | ::: | ::: | Water inlet tubing folded or obstructed|Check that the inlet pipe is not damaged.| | ||
+ | | ::: |- Power light ON \\ - all control light indicators OFF|Fuse in the controller box broken | Replace fuse | | ||
+ | |reduced flow of purified water|- Power light ON \\ - indicator PUMP light ON \\ - Pump running continuously|Pre-treatment clogged| Replace the pre-treatment filters before the RO membrane| | ||
+ | | ::: | ::: | Weak flow of inlet water|Increase the flow of the inlet water| | ||
+ | | ::: | ::: | Reverse osmosis membrane clogged|Change the membrane | | ||
+ | | ::: |Pump leakage | Pump damaged|Replace the pump | | ||
+ | | ::: | ::: | Flow restrictor blocked| Replace flow restrictor | | ||
+ | | ::: | - Power light ON \\ - indicator PUMP light ON \\ - Pump always OFF|Pump fuse burned |Replace the pump fuse | | ||
+ | | Starts and stops continuously | Light indicators FULL and PUMP alternate| Defective high-pressure sensor or check valve|Change pressure sensor and check valve| | ||
+ | |Lack of pressure on purified water output| - Power light ON | Watertank hand valve closed|Turn ON the watertank hand valve| | ||
+ | | ::: | ::: | Requested pure water volume above the capacity of the water purifier|Wait the watertank is filled again\\ | | ||
+ | | ::: | ::: | Lack of pressure in the pressurised watertank|Re-adjust the air-pressure of the water tank| | ||
+ | |purified water conductivity too high| |Resin saturated| Replace resin cartridge or bottle| | ||
+ | |Drain flowing continuously| - POWER light ON \\ - FULL indicator ON \\ - Pump indicator OFF | Inlet solenoid valve defective|Replace the inlet solenoid valve| | ||
+ | | ::: | - Power light ON \\ - Pump indicator ON \\ - Auto-Flush not activated | Flush solenoid valve defective|Replace the flush solenoid valve| | ||
+ | | Conductivity meter display OFF| | Batteries out of order| Replace the batteries| | ||
+ | |||
===== Q & A ===== | ===== Q & A ===== | ||
+ | * [[waterclasses|What are the different qualities of water ?]] | ||
* [[osmosisandresin|What's the difference between RO and DI water purification?]] | * [[osmosisandresin|What's the difference between RO and DI water purification?]] | ||
* [[carbon|Why carbon to filter water ?]] | * [[carbon|Why carbon to filter water ?]] | ||
Line 48: | Line 72: | ||
* [[whereUV|Is the UV sterilizer to be connected before or after the water purifier?]] | * [[whereUV|Is the UV sterilizer to be connected before or after the water purifier?]] | ||
- | |||
- | ===== Troubleshooting ===== | ||
- | [[ROtroubleshooting|troubleshooting]] | ||