Line 38: Line 38:
 **This manual process is similar to the flush function available on the system equipped with an electronic controller, which allows to rince the membrane easily just by pressing a button.** **This manual process is similar to the flush function available on the system equipped with an electronic controller, which allows to rince the membrane easily just by pressing a button.**
 </​WRAP>​ </​WRAP>​
 +
 +===== Troubleshooting =====
 +|**PROBLEM**|**CONDITION**|**POTENTIAL CAUSES **|**SUGGESTED ACTION **|
 +| Water purifier OFF| Power light OFF|No power supply|- Verify the power supply wires \\ - Switch power ON |
 +| ::: | ::: | Power supply fuse is broken| Check and replace the fuse|
 +|No production of purified water|- Power light  ON  \\ - indicator PUMP is off \\ - indicator SOURCE is ON| The inlet tap water supply is closed or has insufficient pressure|Improve water supply|
 +| ::: | ::: | Water inlet tubing folded or obstructed|Check that the inlet pipe is not damaged.|
 +| ::: |- Power light  ON  \\ - all control light indicators OFF|Fuse in the controller box broken | Replace fuse |
 +|reduced flow of purified water|- Power light  ON  \\ - indicator PUMP light ON  \\ - Pump running continuously|Pre-treatment clogged| Replace the pre-treatment filters before the RO membrane|
 +| ::: |  ::: | Weak flow of inlet water|Increase the flow of the inlet water|
 +| ::: | ::: | Reverse osmosis membrane clogged|Change the membrane |
 +| ::: |Pump leakage | Pump damaged|Replace the pump |
 +| ::: | ::: | Flow restrictor blocked| Replace flow restrictor |
 +| ::: | - Power light ON \\ - indicator PUMP light ON \\ - Pump always OFF|Pump fuse burned |Replace the pump fuse |
 +| Starts and stops continuously | Light indicators FULL and PUMP alternate| Defective high-pressure sensor or check valve|Change pressure sensor and check valve|
 +|Lack of pressure on purified water output| - Power light ON  | Watertank hand valve closed|Turn ON the watertank hand valve|
 +| ::: | ::: | Requested pure water volume above the capacity of the water purifier|Wait the watertank is filled again\\ |
 +| ::: | ::: | Lack of pressure in the pressurised watertank|Re-adjust the air-pressure of the water tank|
 +|purified water conductivity too high|   ​|Resin saturated| Replace resin cartridge or bottle|
 +|Drain flowing continuously| - POWER light ON  \\ - FULL indicator ON  \\ - Pump indicator OFF   | Inlet solenoid valve defective|Replace the inlet solenoid valve|
 +| ::: | - Power light ON   \\ - Pump indicator ON \\ - Auto-Flush not activated | Flush solenoid valve defective|Replace the flush solenoid valve|
 +| Conductivity meter display OFF|  | Batteries out of order| Replace the batteries|
 +
 ===== Q & A ===== ===== Q & A =====
 +  * [[waterclasses|What are the different qualities of water ?]]
   * [[osmosisandresin|What'​s the difference between RO and DI water purification?​]]   * [[osmosisandresin|What'​s the difference between RO and DI water purification?​]]
   * [[carbon|Why carbon to filter water ?]]   * [[carbon|Why carbon to filter water ?]]
Line 47: Line 71:
   * [[whenreplacingmembrane|When to replace the reverse osmosis membrane?]]   * [[whenreplacingmembrane|When to replace the reverse osmosis membrane?]]
   * [[whereUV|Is the UV sterilizer to be connected before or after the water purifier?]]   * [[whereUV|Is the UV sterilizer to be connected before or after the water purifier?]]
- 
- 
-===== Troubleshooting ===== 
-[[ROtroubleshooting|troubleshooting]] 
-|**PROBLEMS**|**POSSIBLES CAUSES **|**SUGGESTED ACTION **| 
-| Power light  OFF|No power supply|Verify the power supply wires ; switch power ON | 
-| ::: |On power supply fuse is broken| Check and replace the fuse| 
-|   - Power light  ON  - No production of purified water   - indicator PUMP is off  - indicator SOURCE is ON| The inlet tap water supply is closed or has insufficient pressure|Improve water supply| 
-| ::: | Water inlet tubing folded or obstructed|Check that the inlet pipe is not damaged.| 
-|   - Power light  ON  - No production of purified water  - all control light indicators OFF|A fuse is broken :\\ - pump fuse \\ - fuse in the controller box| Check and replace the fuse| 
-|   - Power light  ON  - No or low purified water production ​ - indicator PUMP light ON  - Pump running|Pre-treatment clogged| Replace the pre-treatment filters before the RO membrane| 
-| ::: | Weak flow of inlet water|Increase the flow of the inlet water| 
-| ::: |Reverse osmosis membrane clogged|Change the membrane | 
-|   - Power light ON  - No or low purified water production ​ - indicator PUMP light ON  - Pump always OFF|Pump fuse burned |Replace the pump fuse | 
-| Light indicators FULL and PUMP alternate|Defective high-pressure sensor or check valve|Change pressure sensor and check valve| 
-|   - Power light ON  - Lack of pressure and low  flow output| Watertank hand valve closed|Turn ON the watertank hand valve| 
-| ::: |Requested pure water volume above the capacity of the water purifier|Wait the watertank is filled again\\ | 
-| ::: |Lack of pressure in the pressurised watertank|Re-adjust the air-pressure of the water tank| 
-| purified water conductivity too high|Resin saturated| Replace resin cartridge or bottle| 
-|   - POWER light ON  - FULL indicator ON  - Pump OFF  - Drain flowing continuously | Inlet solenoid valve defective|Replace the inlet solenoid valve| 
-| Conductivity meter display OFF| Batteries out of order| Replace the batteries| 
- 
- 
  
  