| - | ====== OSMOSERS-TROUBLESHOOTING2 ====== | ||
| - | <sub>{{OSMOSERS-TROUBLESHOOTING2.odt|Original file}}</sub> | ||
| - | |||
| - | |**PROBLEMS**|**POSSIBLES CAUSES **|**SUGGESTED ACTION **| | ||
| - | | __Power light O____FF__|No power supply|Verify the power supply wires ; switch power ON | | ||
| - | | ::: |On power supply fuse is broken| Check and replace the fuse| | ||
| - | | - __Power light O____N__  - No production of purified water - indicator PUMP is off - indicator SOURCE is ON| The inlet tap water supply is closed or has insufficient pressure|Improve water supply| | ||
| - | | ::: | Water inlet tubing folded or obstructed|Check that the inlet pipe is not damaged.| | ||
| - | | - __Power light O____N__  - No production of purified water - __Power light ON__ - all control light __indicators OFF__|A fuse is broken :\\ - pump fuse \\ - fuse in the controller box| Check and replace the fuse| | ||
| - | | - __Power light O____N__  - No or low purified water production  - indicator PUMP light ON - Pump running|  Pre-treatment clogged| Replace the pre-treatment filters before the RO membrane| | ||
| - | | ::: | Weak flow of inlet water|Increase the flow of the inlet water| | ||
| - | | ::: |Reverse osmosis membrane clogged|Change the membrane | | ||
| - | | - __Power light O____N__  - No or low purified water production  - indicator PUMP light ON - Pump always OFF|Pump fuse burned |Replace the pump fuse | | ||
| - | | Light indicators FULL and PUMP alternate|Defective high-pressure sensor or check valve|Change pressure sensor and check valve| | ||
| - | | - __Power light OFF__ - Lack of pressure and low flow output| Watertank hand valve closed|Turn ON the watertank hand valve| | ||
| - | | ::: |Requested pure water volume above the capacity of the water purifier|Wait the watertank is filled again\\ | | ||
| - | | ::: |Lack of pressure in the pressurised watertank|Re-adjust the air-pressure of the water tank| | ||
| - | | purified water conductivity too high|Resin saturated| Replace resin cartridge or bottle| | ||
| - | | - POWER light ON - FULL indicator ON - Pump OFF - Drain flowing continuously | Inlet solenoid valve defective|Replace the inlet solenoid valve| | ||
| - | | Conductivity meter display OFF| Batteries out of order| Replace the batteries| | ||
| - | |||
| - | |||