-====== OSMOSERS-TROUBLESHOOTING ====== 
-<​sub>​{{OSMOSERS-TROUBLESHOOTING.odt|Original file}}</​sub>​ 
- 
-|**PROBLEMS**|**POSSIBLES CAUSES **|**SUGGESTED ACTION **| 
-| Power light  OFF|No power supply|Verify the power supply wires ; switch power ON | 
-| ::: |On power supply fuse is broken| Check and replace the fuse| 
-|   - Power light  ON  - No production of purified water   - indicator PUMP is off  - indicator SOURCE is ON| The inlet tap water supply is closed or has insufficient pressure|Improve water supply| 
-| ::: | Water inlet tubing folded or obstructed|Check that the inlet pipe is not damaged.| 
-|   - Power light  ON  - No production of purified water  - all control light indicators OFF|A fuse is broken :\\ - pump fuse \\ - fuse in the controller box| Check and replace the fuse| 
-|   - Power light  ON  - No or low purified water production ​ - indicator PUMP light ON  - Pump running|Pre-treatment clogged| Replace the pre-treatment filters before the RO membrane| 
-| ::: | Weak flow of inlet water|Increase the flow of the inlet water| 
-| ::: |Reverse osmosis membrane clogged|Change the membrane | 
-|   - Power light ON  - No or low purified water production ​ - indicator PUMP light ON  - Pump always OFF|Pump fuse burned |Replace the pump fuse | 
-| Light indicators FULL and PUMP alternate|Defective high-pressure sensor or check valve|Change pressure sensor and check valve| 
-|   - Power light ON  - Lack of pressure and low  flow output| Watertank hand valve closed|Turn ON the watertank hand valve| 
-| ::: |Requested pure water volume above the capacity of the water purifier|Wait the watertank is filled again\\ | 
-| ::: |Lack of pressure in the pressurised watertank|Re-adjust the air-pressure of the water tank| 
-| purified water conductivity too high|Resin saturated| Replace resin cartridge or bottle| 
-|   - POWER light ON  - FULL indicator ON  - Pump OFF  - Drain flowing continuously | Inlet solenoid valve defective|Replace the inlet solenoid valve| 
-| Conductivity meter display OFF| Batteries out of order| Replace the batteries| 
- 
- 
  