PROBLEMS | POSSIBLES CAUSES | SUGGESTED ACTION | ||
- No production of purified water | indicator PUMP switched off | - Power switch is on OFF position. | Switch ON the device | |
- The water supply of the tap closed or insufficient pressure (in that case the light “SOURCE” is on). | Check the water supply | |||
- water hose folded or obstructed | Check that the inlet pipe is not damaged. | |||
- No production of purified water - Power light ON - all control light indicators OFF | On fuse is broken : - on power supply - in the controller box | Check and replace the fuses | ||
- Not or low purified water production - indicator PUMP light ON - Pump running | Pre-treatment clogged | Replace the pre-treatment filters before the RO membrane | ||
Low flow of inlet water | Increase the flow of the inlet water | |||
Reverse osmosis membrane clogged | Change the membrane | |||
- Not or low purified water production - indicator PUMP light ON - Pump always OFF | Pump fuse burned | Replace the pump fuse | ||
Light indicators FULL and PUMP alternate | Defectiveness of the high-pressure sensor or check valve. | Pressure sensor and check valve | ||
Lack of pressure and low flow output | - Watertank hand valve closed. | Turn ON the watertank hand valve | ||
- volume of pure water need above the specification of the water purifier | Wait the watertank is filled again | |||
- Lack of pressure in the pressurised watertank | Re-adjust the air-pressure of the water tank | |||
purified water conductivity too high | Resin saturated. | Replace resin cartridge or bottle | ||
- POWER & FULL indicators ON - Pump OFF - Drain flowing continuously | - Inlet solenoid valve defective | Replace the inlet solenoid valve | ||
Conductivity meter display OFF | Batteries out of order | Replace the batteries |