This is an old revision of the document!
PROBLEMS | POSSIBLES CAUSES | SUGGESTED ACTION | |
- No production of purified water | indicator PUMP switched off | - Power switch is on OFF position. | Switch ON the device |
- The water supply of the tap closed or insufficient pressure (in that case the light “SOURCE” is on). | Check the water supply | ||
- water hose folded or obstructed | Check that the inlet pipe is not damaged. | ||
- No production of purified water - Power light ON - all control light indicators OFF | On fuse is broken : - on power supply - in the controller box | Check and replace the fuses | |
- Not or low purified water production - indicator PUMP light ON - Pump running | Pre-treatment clogged | Replace the pre-treatment filters before the RO membrane | |
Low flow of inlet water | Increase the flow of the inlet water | ||
Reverse osmosis membrane clogged | Change the membrane | ||
- Not or low purified water production - indicator PUMP light ON - Pump always OFF | Pump fuse burned | Replace the pump fuse | |
Light indicators FULL and PUMP alternate | Defectiveness of the high-pressure sensor or check valve. | Pressure sensor and check valve | |
Lack of pressure and low flow output | - Watertank hand valve closed. | Turn ON the watertank hand valve | |
- volume of pure water need above the specification of the water purifier | Wait the watertank is filled again | ||
- Lack of pressure in the pressurised watertank | Re-adjust the air-pressure of the water tank | ||
purified water conductivity too high | Resin saturated. | Replace resin cartridge or bottle | |
- POWER & FULL indicators ON - Pump OFF - Drain flowing continuously | - Inlet solenoid valve defective | Replace the inlet solenoid valve | |
Conductivity meter display OFF | Batteries out of order | Replace the batteries |