OSMOSERS-TROUBLESHOOTING

PROBLEMS POSSIBLES CAUSES SUGGESTED ACTION
- No production of purified water indicator PUMP switched off - Power switch is on OFF position. Switch ON the device
- The water supply of the tap closed or insufficient pressure (in that case the light “SOURCE” is on).Check the water supply
- water hose folded or obstructed Check that the inlet pipe is not damaged.
- No production of purified water - Power light ON - all control light indicators OFF On fuse is broken :
- on power supply
- in the controller box
Check and replace the fuses
- Not or low purified water production - indicator PUMP light ON - Pump runningPre-treatment cloggedReplace the pre-treatment filters before the RO membrane
Low flow of inlet waterIncrease the flow of the inlet water
Reverse osmosis membrane cloggedChange the membrane
- Not or low purified water production - indicator PUMP light ON - Pump always OFFPump fuse burned Replace the pump fuse
Light indicators FULL and PUMP alternateDefectiveness of the high-pressure sensor or check valve.Pressure sensor and check valve
Lack of pressure and low flow output- Watertank hand valve closed.Turn ON the watertank hand valve
- volume of pure water need above the specification of the water purifierWait the watertank is filled again
- Lack of pressure in the pressurised watertankRe-adjust the air-pressure of the water tank
purified water conductivity too highResin saturated.Replace resin cartridge or bottle
- POWER & FULL indicators ON - Pump OFF - Drain flowing continuously - Inlet solenoid valve defectiveReplace the inlet solenoid valve
Conductivity meter display OFF Batteries out of order Replace the batteries