OSMOSERS-TROUBLESHOOTING2

Original file

PROBLEMSPOSSIBLES CAUSES SUGGESTED ACTION
Power light OFFNo power supplyVerify the power supply wires ; switch power ON
On power supply fuse is broken Check and replace the fuse
- Power light ON - No production of purified water - indicator PUMP is off - indicator SOURCE is ON The inlet tap water supply is closed or has insufficient pressureImprove water supply
Water inlet tubing folded or obstructedCheck that the inlet pipe is not damaged.
- Power light ON - No production of purified water - Power light ON - all control light indicators OFFA fuse is broken :
- pump fuse
- fuse in the controller box
Check and replace the fuse
- Power light ON - No or low purified water production - indicator PUMP light ON - Pump running Pre-treatment clogged Replace the pre-treatment filters before the RO membrane
Weak flow of inlet waterIncrease the flow of the inlet water
Reverse osmosis membrane cloggedChange the membrane
- Power light ON - No or low purified water production - indicator PUMP light ON - Pump always OFFPump fuse burned Replace the pump fuse
Light indicators FULL and PUMP alternateDefective high-pressure sensor or check valveChange pressure sensor and check valve
- Power light OFF - Lack of pressure and low flow output Watertank hand valve closedTurn ON the watertank hand valve
Requested pure water volume above the capacity of the water purifierWait the watertank is filled again
Lack of pressure in the pressurised watertankRe-adjust the air-pressure of the water tank
purified water conductivity too highResin saturated Replace resin cartridge or bottle
- POWER light ON - FULL indicator ON - Pump OFF - Drain flowing continuously Inlet solenoid valve defectiveReplace the inlet solenoid valve
Conductivity meter display OFF Batteries out of order Replace the batteries