User Tools

Site Tools


whenreplacingmembrane

This is an old revision of the document!


Membrane replacement test

  • Measure the conductivity after the reverse osmosis membrane
  • Measure the conductivity of the tap water
  • Do the following calculation : 100 * (conductivity reverse osmosis) / (conductivity raw water)
  • If the value is less than 10%, the membrane is still usable.

In the other case, the reverse osmosis membrane begins to saturate.

whenreplacingmembrane.1553681532.txt.gz · Last modified: 2019/03/27 10:12 by thierry