User Tools

Site Tools


start

This is an old revision of the document!


Introduction

The water purifiers proposed by DIASYS are a combination of 2 elements:

  • a reverse osmosis system removing 95% of impurities
  • a deionising filter producing pure water(conductivity below 5µS/cm)

The reverse osmosis membrane has the advantage to prolong the lifetime of the deionising filter.

The water purifiers designed by Diasys are especially built for providing deionised water to clinical chemistry and immunology analysers.

catalog

A reverse osmosis system works well with water at ambiant temperature. The tap water should not be too hot (above 35°C) or too cold ( below 10°C). Outside these bounds, the membrane might be damaged irreversibly. The optimum production of water is obtained at 25°C.

In order to activate the reverse osmosis process, the tap water pressure in the membrane must be strong enough, above 2.5 BAR ( ideal pressure is generally around 4.5 BAR) but can vary depending of the membrane. When the tap water pressure is too low or vary, a booster pump is required. It is recommended to rince the membrane every month using the following process:

  • disconnect the restrictor
  • leave water flowing during 20 minutes
  • reconnect the restrictor

This manual process is similar to the flush function available on the system equipped with an electronic controller, which allows to rince the membrane easily just by pressing a button.

Troubleshooting

PROBLEMSPOTENTIAL CAUSES SUGGESTED ACTION
Power light OFFNo power supplyVerify the power supply wires ; switch power ON
Power supply fuse is broken Check and replace the fuse
- Power light ON - No production of purified water - indicator PUMP is off - indicator SOURCE is ON The inlet tap water supply is closed or has insufficient pressureImprove water supply
Water inlet tubing folded or obstructedCheck that the inlet pipe is not damaged.
- Power light ON
- No production of purified water
- all control light indicators OFF
A fuse is broken :
- pump fuse
- fuse in the controller box
Check and replace the fuse
- Power light ON - No or low purified water production - indicator PUMP light ON - Pump runningPre-treatment clogged Replace the pre-treatment filters before the RO membrane
Weak flow of inlet waterIncrease the flow of the inlet water
Reverse osmosis membrane cloggedChange the membrane
- Power light ON - No or low purified water production - indicator PUMP light ON - Pump always OFFPump fuse burned Replace the pump fuse
Light indicators FULL and PUMP alternateDefective high-pressure sensor or check valveChange pressure sensor and check valve
- Power light ON - Lack of pressure and low flow output Watertank hand valve closedTurn ON the watertank hand valve
Requested pure water volume above the capacity of the water purifierWait the watertank is filled again
Lack of pressure in the pressurised watertankRe-adjust the air-pressure of the water tank
purified water conductivity too highResin saturated Replace resin cartridge or bottle
- POWER light ON - FULL indicator ON - Pump OFF - Drain flowing continuously Inlet solenoid valve defectiveReplace the inlet solenoid valve
Conductivity meter display OFF Batteries out of order Replace the batteries

Q & A